4027A60M, 45-65 MHz Precision Power Sensor (CALDATA included)

Call (866) 695-4569 for Quote or Email us for more information
45-65 MHz, 3 W to 6 kW
4027A60M, RF Power Sensor is designed for use in semiconductor processing and other precision process applications. At the calibrated frequency and power level, these sensors are capable of 1% accuracy. Intended for use with Bird 4421A Series Precision Multifunction Power Meters, these sensors provide a threefold improvement in long term unit to unit accuracy.
Problems/Solutions
- Poor production yields - 1% accuracy at specified frequencies and power levels
- Lack of confidence in measurements - Calibration traceable to NIST
- Complex tools requiring calibration each time
- Plug and Play with 4421A Meter
- Unit does not need to be field calibrated before use
- Calibrate only once every six months
- Harmonic content interfering with measurements
- Wide range of applications requiring various input and output connectors - Dozens of connector options available
Specifications:
- Frequency Range: 45-65 MHz
- Power Range: 3 W to 6 kW
- Accuracy: ±1% at calibration frequencies and power levels
±2 % at other frequency and power levels
Add 2% to uncertainty outside 25 ± 10 °C - Calibration Power Level: 1000W units: 700 watts
10kW units: 1700 watts - Uniformity: 2 % maximum unit to unit, at calibration frequency and power levels
- Speed: 2 readings per second
- Maximum Power: 10 kW units - 12 kW (max)
1 kW units - 1.2 kW (max) - Harmonic Content: -50 dBc (max)
- VSWR Range: 1.0-2.0
- Directivity: 28 dB
- Insertion Loss: <0.05 dB
- External DC: 12 VDC, supplied from 4421A Multifunction Power Meter
- Connectors: Input and Output Connectors MUST be ordered separately at additional cost
- Calibration Cycle: 6 month. Performance before and performance after data to be supplied for units